My Cart: 0 item(s)

Product Search

Secure Checkout
Sutter-Lambda DG 4 Plus Ultra High Speed Wavelength Switcher
    Sutter-Lambda DG 4 Plus Ultra High Speed Wavelength Switcher
    Purchase Sutter-Lambda DG 4 Plus Ultra High Speed Wavelength Switcher
    • Sutter-Lambda DG 4 Plus Ultra High Speed Wavelength Switcher

      SKU: DG-4PLUS
      Product Code: sutter-lambda-dg-4plus
      Vendor: Sutter Instrument
    • Price: $22,615.00

      Quantity in Basket:  None

      Product Documents

    •  

      Do it Now

      Fill out the rest of this form, and hit "submit." A member of our sales team will get back to you shortly.

      *Fields in bold are required.*


    Sutter-Lambda DG 4 Plus Ultra High Speed Wavelength Switcher


    The Lambda DG-4/DG-5 Plus is a complete illumination system offering speed and versatility for experiments requiring rapid wavelength switching. New digital servo technology allows faster filter switching and 30% greater light output than the first generation DG-4/DG-5. The instrument retains all the advantages of interference filter based systems, yet eliminates the temporal constraints imposed by traditional filter changing devices like filter wheels. Switching between any two wavelengths is achieved in 0.5msec, allowing the user to perform real-time video imaging. The high switching speed of the Lambda DG-4/DG-5 Plus facilitates the ability to follow fast changes in ion concentrations in dual wavelength ratio imaging applications and to monitor other concomitant changes in the studied system at additional wavelengths.


    Narrow bandpass systems, such as single cavity interference filters, grating monochromators, and A.O. modulators, pass unwanted harmonics of the desired wavelength. Also, with variable wavelength devices, it is not always possible to obtain sufficient blocking of out-of-band wavelengths. Modern interference filters, as used in the Lambda DG-4/DG-5 Plus, do not pass harmonics and have integral blocking characteristics 1000 times better than typical monochromator systems. For the same full width at half maximum (FWHM), interference filters have a narrower spectral bandpass due to the absence of the slit function.


    The dual galvanometer design of the Lambda DG-4/DG-5 Plus allows tuning of the relative intensities at each wavelength. This adjustment, in variable wavelength devices with a single optical path, is difficult to obtain. Unlike monochromators and other wavelength selective systems, the Lambda DG-4/DG-5 Plus can be used as a source of white light when required.


    How it works:
    This unique optical design of the Lambda DG-4/DG-5 Plus is based on dual scanning galvanometers utilizing interference filters for wavelength selection. The light from the xenon arc lamp is focused on the first galvanometer mirror which directs it via a parabolic mirror, through one of the interference filter channels. Following the filter, a second parabolic and a second galvanometer mirror collect and redirect the light into the light guide. A cold mirror in the beginning of the light path eliminates the IR radiation, reducing significantly the amount of heat absorbed by the optics and the sample. Cold mirrors modified to pass near-IR to 780 to 880nm are also available.


    A built-in shutter function allows reduction of light intesity by five to six orders of magnitude. For applications requiring mechanical shuttering, like time lapse experiments or very sensitive samples, the SmartShutter® can be installed in the device to assure that the light is completely blocked. A newly integrated SmartShutter controller eliminates the need to purchase a separate control device.


    The standard system, the Lambda DG-4 Plus, holds up to four 25mm interference filters. The Lambda DG-5 Plus, a five filter version, accommodates three 18mm and two 25mm filters. While the switching time between any two wavelengths is done in less than 0.5msec, the dwell time at any wavelength is arbitrarily set by the user.


    The liquid light guide can be coupled to the illumination port of most microscopes using an adapter which can be purchased separately. Please refer to the Microscope section for further information. Extended output ranges are possible with various cold mirror and light guide combinations. Phone Sutter to discuss your specific application requirements.


      FEATURES
    • NEW: USB support
    • NEW: Built-in SmartShutter® control
    • NEW: Expanded menu options and enhanced triggering
    • Integral shuttering
    • Integral neutral density filtering
    • Two outputs for monitoring filter position
    • Turbo Blanking
    • Direct computer control via parallel, serial, or USB interface
    • Switches in 0.5msec

      COMMON APPLICATIONS
    • Fluorescent microscopy
    • Calcium imaging
    • FURA
    • Optogenetics
    • Ultra high speed wavelength selection

    • TECHNICAL SPECIFICATIONS


      Lamp Type
      300 Watt pre-aligned, ozone free xenon arc bulb
      (full spectrum available upon request)

      Output Range
      330nm to 650nm - ozone free
      300nm to 650nm - full spectrum

      Lamp Lifetime
      1,000 hours
      (Bulb has a prorated warranty for 500 hours. Longer life depends on application. Expected life is 1000 hours)

      Power Consumption
      350 Watts

      Filter Diameter
      DG-4: Four, 25mm (1inch) 
      DG-5: Two, 25mm (1inch) and Three, 18mm

      Dimensions
      10in x 10in x 19in | 25cm x 25cm x 48cm

      Light Guide
      2 meters long
      3mm diameter

      Weight
      45lbs | 20kg

      Electrical
      115/230 Volts
      50/60 Hertz power line

      RoHS Compliant