    BrightLine® single-band λ/2 flatness super-resolution laser filter set, optimized for 405 nm laser sources

    This set is designed to take advantage of the superior resolution, sensitivity, and image fidelity of today's state-of-the-art laser-based microscopes. The filter spectra are optimized to capture the maximum fluorescence from DAPI and other similar fluorophores while offering superb blocking of popular 405 nm or 375 nm solid-state laser diodes. The dichroic offers industry-leading λ/2 flatness for minimal focus shift and aberrations of the laser beam spot up to 10 mm diameter to enable popular imaging and Super-Resolution techniques including TIRF, PALM, STORM, and Structured-Illumination.

     

    Optical Specifications
    Specification Value
    Blocking BrightLine filters have blocking far exceeding OD 6 as needed to ensure a black background, even when using modern low-noise CCD cameras. The blocking is optimized for microscopy applications using our exclusive SpecMaker™ fluorescence filter design.

    General Filter Specifications
    Specification Value
    Microscope Compatibility BrightLine filters are available to fit Leica, Nikon, Olympus, Zeiss, and Aperio microscopes.
    Optical Damage Rating Testing has proven to show no signs of degradation when exposed to at least 6.0 W of power from an unfiltered xenon arc lamp over a 25 mm diameter (corresponding to 1.2 W/cm2) for over 500 hours.

    Physical Filter Specifications
    Specification Value
    Scratch-Dig 60-40
    Exciter/Emitter Orientation Arrow on ring indicates preferred direction of propagation of light.
    Dichroic Orientation "Reflective coating side" should face toward light source and sample.
    Flatness Loose and cube mounted filter sets deliver certified dichroic flatness ≤ λ/2 PV per inch @ 632.8 nm