The
heart of the BIONOMIC SYSTEM© is the BC-100 CONTROLLER
that regulates the temperature of a variety of "cells"
designed to accept conventional substrate holder such as 60
and 35mm petri dishes, microscope slides, Dvorak chambers.
Open and closed perfusion chambers are available. The flexibility
of the system design allows special designed cells to be supplied
for unique applications.
All Bionomic Cells may be adapted with a gas purging system
that protects the specimen from the outside atmosphere.
Two gases such as N2 and CO2 may be accurately mixed in any
ratio and gently purged over the substrate. The gas may be
heated to reduce condensation when the Bionomic Cell and specimen
are at sub-ambient temperature.
Bionomic Cells designed for low temperatures (less than 12°
C) are water cooled to remove excess heat from the cooling
elements. A miniature, closed heat exchanger, the HEC-400
eliminates the need of using tap water and drains, preventing
any possibility of water spills or floods and is recommended
for use with the water cooled cells.
|
Features
- Reaches
"SET" Temperature in Seconds
-
Computer Compatible for Temperature Set and Programmability
-
EMF-Free for Sub-Nanovolt Probing
-
Atmosphere Control with Gas Mixing
-
Unobstructed Optical path for Short Working Distance Objectives
-
Open Design of Cells for Probe and Micromanipulator Access
-
Digital Read Out of Cell Temperature at All Times
|
|
Technical
Specifications
| Temperature
Control Range: |
-10°
to 60° C at the press of a button |
| Temperature
Accuracy: |
±
0.2° C |
| Temperature
Stability: |
±
0.1° C |
*Specifications subject to change without notice. |