Technical
Specifications: |
|
|
Injectors |
- Split / Splitless supplied as standard
- Min Temp 60C
- Max Temp 280C
- Split Linearity better than 3% up to 200:1
|
|
|
Detectors |
- Choice of FID, TCD or ECD
- Temperature range from 100C to 300C
|
|
|
FID |
- FID minimum detectable quantity is 5 x 10^-12 grams of carbon (signal 3x noise level)
- FID linear range is 10^6
|
|
|
TCD |
- TCD has four tungsten filaments
- TCD filament temperatures are 180C, 240C or 320C
|
|
|
ECD |
- ECD minimum detectable quantity is 0.05 x 10^-12 grams of lindane (signal 3x noise level)
- ECD linear range is 10^3
|
|
|
Capillary On Column Adapters |
Available as an option accessory |
| |
|
|
- Electronic Pressure control of carrier gas of 0.3 to 50Psi fully programmable with up to five ramps. Maximum total flow 250ml/min
- Detector gas control fixed sinter (apply the specified pressure to the instrument and the detector will work)
- Optional two channel gas box to allow variability of detector gasses
|
|
|
Instrument Control |
- keypad, with 2 x 16 character backlit LCD
- 5 stored methods
- Standard RS232C interface allows methods to be stored and loaded from pc using Chromstar GC control Software (optional accessory)
- 1V/10V analogue outputs
- Run line control allows the use of most recorder, integrators and PC based integration packages
|
|
|
Gas Sampling Valves |
Available as an option accessory |
|
|
|
- Accommodates all capillary columns up to a 7" cage, max length 120m (5"cages required for above 60m)
- Packed Column capability available as an optional accessory
- Temperature range is from 15C above ambient to 300C
- 5 ramp temperature programming
- Ramp rates are up to 40C/min below 200C, and up to 20C/min above 200C
- Max isothermal temperature 240C
|
|
|
Size |
(H) 16cm x (W) 41cm x (D) 34cm |
|
|
Power |
800VA, 230V/50/60Hz, 115V/50/60Hz |
|
|
Weight |
7.5Kg |
*Specifications subject to change without notice. |